Icon name
Icon
Inj. Delay (s)
Type
Ratio (%)
Pressure Limit
Contrast
Saline
Get Injector Parame‐
ters
Add Phase
Delete Phase
Apply
Cancel
NOTE:
There is no Xtream Injector button for Scout scans.
328
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
Description
Injection delay can be set in increments of seconds.
•
0 no delay prior to the injection of this phase.
•
1 to max — injection delay of the injector in sec‐
onds prior to the start of this phase. This is a timed
pause.
•
Hold — the injection for this phase is started from
the injector by pressing the start button. This is on‐
ly valid for phases two through five.
Select the type of injection: Saline, Contrast or Contrast-
Saline.
Selecting Contrast-Saline opens the Ratio fields.
Contrast-Saline is selected in Type .
Available only when
The upper frame is for contrast, the lower frame if for
Saline. The range is 5:95 and 95:5.
When mixed phase is set from CT user interface, the in‐
Injector Monitor is dismissed. If you
dication of ratio on
want to set the ratio on Injector Monitor, those buttons
are displayed.
The pressure limit value sets the limit at which when
reached will stop the injection.
The system displays the remaining volume of contrast
media and saline.
Selecting [Get Injector Parameters] retrieves the values
currently set on the injector and displays them in the In‐
jector Settings popover. Click [Apply] to choose these
values or click [Cancel] to return to the original values.
Adds a phase.
Deletes a phase.
Clicking [Apply] when the injector is armed, disarms the
injector, but the parameters will NOT be updated on the
injector. Open the popover again and click [Apply] to up‐
date the parameters.
Parameter values are not updated and remain at the
values present when the pop up window is opened.