7 Build Protocols with Notification Values
Use this procedure to set Notification Values (NV) in user protocols.
NV values can only be set in Protocol Management.
1. From the image monitor, click
Mode menu, click Protocol Management , then Protocol Management .
2. From the
CT Login authentication window will open.
The
3. Enter your Log On Name and Password and click [OK] to continue.
Protocol Selection window, select Adult or Pediatric .
4. From the
Anatomical area, select an anatomical region.
5. From the
Protocol Selection window, select a protocol and click [Edit].
6. From the
7. Click the series task to go to the first non-scout series in the protocol.
Dose Display area, enter the CTDI
8. From the
9. Click [Confirm Changes].
10. Click [Save] or [Save As] to save the new protocol parameters.
○
The system compares all the CTDI
NV value.
○
The protocol parameters will not be accepted until the NV value is no longer exceeded
or the user has entered a
parameters exceeding the NV value.
○
Saving a protocol with a dose level exceeding an NV demands a high degree of
consideration. In most cases, you will click [Cancel] to go back to adjust the parameters.
If there is a clinical reason to save a protocol exceeding an NV, enter the diagnostic
reason.
Diagnostic Reason field, enter the reason for the scan exceeding the NV value.
11. In the
12. Click [Confirm] to accept the
13. Click [Confirm Changes] to accept the protocol scan parameters.
14. Click [Save] or [Save As] to return to the
NOTE:
Viewing or exporting a Protocol Summary in the Audit Tool should be done after the
protocols are saved, not while editing.
148
.
DLP NV value by group.
or
vol
and DLP values in the scan groups to the entered
vol
Diagnostic Reason for continuing to accept the scan
Diagnostic Reason .
Protocol Selection window.
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
7 Build Protocols with Notification Values