2.2 Apply Product Network Filters
Product Network Filters allow you to place a site firewall in front of the CT scanner. Filters may
be configured to restrict which computers may access the system and which system services
are allowed to be controlled by other devices trying to access the operator console such as ftp,
telnet and rlogin. This access is determined by configuring IP addresses allowed to access the
system. DICOM access to the system can also be configured. The system default is for no
services enabled, no IP addresses authorized and DICOM port 4006 enabled.
Consult your Field Engineer and your IT Department to configure Product Network Filters.