Responder 1000/1100 Version V 1.0 Servicing Instructions 227 487 20 ENG Revision G...
Page 2
Caution: During repairs/service interventions, observe the protective measures against damage due to ESD. • GE Medical Systems Information Technologies GmbH is responsible for the effects on safety, reliability, and performance of the device, only if − assembly operations, extensions, readjustments, modifications, or repairs are carried out...
Service Instructions ────────────────────────────────────────────────────────────────── Contents 1 Documentation and nomenclature of GE Medical Systems instrument part Nos....... 5 1.1 Configuration of Instrument part No..................5 1.2 Configuration of the PCB part No ..................5 1.3 Instrument Status Documentation (nominal status) .............. 6 2 General Overview ........................
Page 4
GE Medical Systems Information Technologies Responder 1000/1100 Page 4/38 Service Instructions ────────────────────────────────────────────────────────────────── Revision History Each page of this manual has the document number followed by a revision letter, located at the bottom line of the page. This letter identifies the manual update level. The latest letter of the alphabet corresponds to the most current revision of the document.
Page 5/38 Service Instructions ────────────────────────────────────────────────────────────────── 1 Documentation and nomenclature of GE Medical Systems Information Technologies instrument part Nos 1.1 Configuration of Instrument part No The instrument part No comprises 8 digits, the first 6 digits determining the instrument type, the last 2 digits the instrument version.
GE Medical Systems Information Technologies Responder 1000/1100 Page 6/38 Service Instructions ────────────────────────────────────────────────────────────────── 303 xxx yy Spare part numbers for PCBs tested especially thoroughly 303 numbers are only given to PCBs where the level of testing applied to 388 PCBs is inadequate for implementation when servicing in the field, or where only a complete set of PCBs can be replaced in the field.
────────────────────────────────────────────────────────────────── 2 General Overview These service instructions describe both the Responder 1000 as well as the Responder 1100. Unless a note appears to the contrary, this description applies to both units. Responder 1000 and Responder 1100 are based on the same hardware platform.
Page 8/38 Service Instructions ────────────────────────────────────────────────────────────────── 2.1 Responder 1000 Button to deliver the shock - together with button 8 AC line power indicator Indicator light is green while the battery is charging (line power operation), it is red when the battery...
GE Medical Systems Information Technologies Responder 1000/1100 Page 9/38 Service Instruction ─────────────────────────────────────────────────────────────────── 2.2 Responder 1100 ECG trigger signal input (synchronized defibrillation) Button to deliver the shock - together with button 12 AC line power indicator Button to enable and disable the synchronized defibrillation mode, with indicator (indicator is illuminated during sync.
GE Medical Systems Information Technologies Responder 1000/1100 Page 10/38 Service Instruction ─────────────────────────────────────────────────────────────────── 2.3 Block Diagram, total Unit 227 487 20 Rev G...
The major mechanical components of the unit are the top and bottom shell. The top shell is the basic element carrying the following sub-assemblies: PCB Responder 1000/1100 Keypad (only Responder 1100) The bottom shell holds the battery module and the capacitor which are linked to the PCB Responder 1000/1100 via cables.
Information Technologies Responder 1000/1100 Page 12/38 Service Instruction ─────────────────────────────────────────────────────────────── ──── 3 Interfaces nterfaces 3.1 Internal Interfaces 3.1 Internal Interfaces Interface signal PCB Responder 1000/1100 Interface signal PCB Responder 1000/1100 connector : connector : BattLoad1 BattLoad1 function: connector to battery module type:...
Page 13
GE Medical Systems Information Technologies Responder 1000/1100 Page 13/38 Service Instruction ─────────────────────────────────────────────────────────────────── type: TRIO MATE 4pol signal name level active meaning SHOTAFOL_ Input key “shock” CHATAFOL_ Input key “charge/shock SYNTAFOL_ Input key “sync” Output Ground connector : PaddConn1 function: connector to “Sternum” Paddle...
Page 14
BAT_PLUS battery + BAT_MINUS battery - connector : BattLoad2 function: connection to PCB Responder 1000/1100 type: crimp connection terminal mini 2000 signal name level active meaning Usec1 +18..30V Input charging voltage or battery voltage Input...
GE Medical Systems Information Technologies Responder 1000/1100 Page 15/38 Service Instruction ─────────────────────────────────────────────────────────────────── 3.2 External Interfaces Only the Responder 1100 is equipped with external interfaces. Sync interface 1 1-Volt ECG signal, analog (see chapter 9 Technical Description) 2 Ground, digital 3 SYNC pulse, digital (see chapter 9 Technical Description)
GE Medical Systems Information Technologies Responder 1000/1100 Page 16/38 Service Instruction ─────────────────────────────────────────────────────────────────── 4 Adjustment and Performance Instruction KOPOS (+) SYNC R506 KEYCONN KONEG (-) HV_TRAFO_PRI BATTLOAD R502 R503 must set must set not set 230V MAINS not set 115V not set...
Page 17
GE Medical Systems Information Technologies Responder 1000/1100 Page 17/38 Service Instruction ─────────────────────────────────────────────────────────────────── Where What What to Where to Where to How much and What else to note? (e.g. PCB) (e.g. PCB) measure with? clip to? turn exact? 30 ± 1s until...
Every three seconds the unit try’s a new restart and test whether the error still appears. 5.2 Test Mode The test mode is activated by inserting a jumper on PCB Responder 1000/1100 (J3 5-6). In the test mode, some LED’s of the energy levels (2J to 30J) will indicate any detected malfunction. The table below lists all error codes and the corresponding explanations.
Page 19
GE Medical Systems Information Technologies Responder 1000/1100 Page 19/38 Service Instruction ─────────────────────────────────────────────────────────────────── Error codes and the corresponding explanations Meaning What to do light key error test keypad and paddles light rotary switch error replace PCB light light high voltage error...
The battery module is located in the bottom shell. With the unit open, first disconnect the plug connection BATTLOAD to the PCB. Then remove the 4 attachment screws. Caution: Use only original GE Medical Systems batteries (see chapter 11, Spare Parts List) Return defective battery module to the factory for proper disposal.
GE Medical Systems Information Technologies Responder 1000/1100 Page 21/38 Service Instruction ─────────────────────────────────────────────────────────────────── 7 Troubleshooting Unit cannot be switched on during line-power operating: Green LED (AC line power indicator) fails to light up and unit cannot be switched on line power cord defective or not connected properly ?
GE Medical Systems Information Technologies Responder 1000/1100 Page 22/38 Service Instruction ─────────────────────────────────────────────────────────────────── 8 Maintenance and Technical Inspection 8.1 Testing Equipment defibrillator-tester standard 1V-ECG signal generator 8.2 Technical Inspection Caution! The unit uses high voltage. The inspection should be referred to independent persons with adequate training and experience.
Switch the unit on with energy selector switch (turn rotary switch to the first position 2J). After power-up and during operation, the Responder 1000/1100 runs automatic self-tests. After power-up, the unit beeps and all LED’s will briefly light up. Afterwards the green “On- LED”...
Page 24
GE Medical Systems Information Technologies Responder 1000/1100 Page 24/38 Service Instruction ─────────────────────────────────────────────────────────────────── 8-12 18-22 28-32 47-53 93-107 185-215 278-322 333-387 Synchronized defibrillation Connect the Sync-cable to the defibrillator. Use a standard 1V-ECG signal generator (connected to Pin 1 and 5) or digital pulse (connected to Pin 3 and 2) to generate signals. Set energy selector switch to any energy setting.
GE Medical Systems Information Technologies Responder 1000/1100 Page 25/38 Service Instruction ─────────────────────────────────────────────────────────────────── 8.3 Safety Analysis Test 8.3.1 General Information The suggested Safety Analysis Test refer to the international Standard IEC 601-1. The tests are generally performed with Safety Testers, on most of them, the measuring circuits according IEC 601 are already implemented.
GE Medical Systems Information Technologies Responder 1000/1100 Page 26/38 Service Instruction ─────────────────────────────────────────────────────────────────── 8.3.3 Enclosure Leakage Current This test is performed to measure leakage current from chassis to ground during normal conditions (N.C.) and single fault conditions (S.F.C.). In any case, the leakage current is measured from any exposed conductive parts to ground, the unit under test has to be switched on and off.
GE Medical Systems Information Technologies Responder 1000/1100 Page 27/38 Service Instruction ─────────────────────────────────────────────────────────────────── 8.3.4 Patient Leakage Current This test performs a leakage current test under single fault conditions (S.F.C.) dependent of domestic power outlet with 115 or 230 V AC as source into the floating inputs.
The “Technical Description” section describes the technical data of the unit valid at the time of printing Operating mode • Responder 1000, non-synchronized defibrillation • Responder 1100, non-synchronized and synchronized defibrillation Energy selection • Selectable energy levels, energy to be delivered into a 50-ohm external resistance (max.
Page 29
GE Medical Systems Information Technologies Responder 1000/1100 Page 29/38 Service Instruction ─────────────────────────────────────────────────────────────────── Discharge circuit • serial oscillating circuit with external resistance (patient) • capacitance 35 µF • inductance 45 mH • equivalent resistance 10.7 ohms, in series with external resistance (patient)
Page 30
GE Medical Systems Information Technologies Responder 1000/1100 Page 30/38 Service Instruction ─────────────────────────────────────────────────────────────────── Digital marker output The trigger signal can be checked on the monitor screen. For this purpose, the Responder 1100 supplies a marker signal. Some patient monitors (e.g. Eagle 3000/4000) use this signal to superimpose the trigger signal on the ECG waveform.)
(5 in.) Weight • model without battery approximately 5.4 kg (11.0 lbs) • model with battery approximately 5.8 kg (12.8 lbs) 10 Spare Part List Responder 1000 Version’s Part Number Description 101 166 01 without Battery 230V-240V 101 166 02...
Page 32
GE Medical Systems Information Technologies Responder 1000/1100 Page 32/38 Service Instruction ─────────────────────────────────────────────────────────────────── Operator's Manual Responder 1000 Part Number Description 227 487 01 Operator’s Manual German 227 487 02 Operator’s Manual English 227 487 03 Operator’s Manual French 227 487 04 Operator’s Manual USA...
Page 33
Label for Housing upper shell Responder 1000 (Logo: marquette) Part Number Description 430 520 13 English, U.S. version Label for Housing upper shell Responder 1000 Minidef PRO (Logo: ESAOTE) Part Number Description 430 520 14 Italian 227 487 20 Rev G...
Page 34
Label for Housing upper shell Responder 1100 (Logo: marquette) Part Number Description 430 520 20 English, U.S. version Label for Housing upper shell Responder 1000 Minidef SYC (Logo: ESAOTE) Part Number Description 430 520 11 Italian Label for Housing lower shell Responder 1000/1100...
Page 35
GE Medical Systems Information Technologies Responder 1000/1100 Page 35/38 Service Instruction ─────────────────────────────────────────────────────────────────── Label for Housing lower shell Responder 1000 Part Number Description 2027555-001 WEEE label 2034056-020 China RoHS label Keypad’s Part Number Description 390 001 79 Keypad for Responder 1000 with Battery...
Page 37
GE Medical Systems Information Technologies Responder 1000/1100 Page 37/38 Service Instruction ─────────────────────────────────────────────────────────────────── Supplies for Responder 1100 Defibrillation external Part Number Description 217 337 01 Defib electrode, external, 217 337 02 Defib electrode, external, 303 439 96 Contact plate for Defib electrodes, 1pcs...
Page 46
Diese Tabelle definiert die gültigen Konfigurationen von : Responder 1000 Version 1 This table defines the valid configurations of : Responder 1000 Version 1 Komponente Sachnummer Index kompatibel (Service) in Variante Sw.Rel. Component Part No. Rev. compatible (Service) in Device sw.
Page 47
388 032 79 - D02 D-79007 Freiburg REVISIONS Not equipped in Version : 43367260 Revision-No Index Date / Name SCHEMATIC 38803279 063764 22.02.00/MSG Date / Name DRAWN 18.12.97/BNH APPROVED 18.12.97/BNH RESPONDER 1000/1100 B. HAGER ISSUED [1997, 12, 2, 3] /home/projects_b/f67260/design...
Page 48
388 032 79 - D03 D-79007 Freiburg REVISIONS Not equipped in Version : F67260 Revision-No Index Date / Name SCHEMATIC 38803279 063764 22.02.00/MSG Date / Name 067395 05.07.01/TYR DRAWN 17.12.97/JFS 17.12.97/JFS APPROVED RESPONDER 1000/1100 ISSUED J. SCHULER [1997, 12, 12, 6] /home/projects_b/f67260/design/power_supply...
Page 49
GE Medical Systems Sheet: 388 032 79 - D04 D-79111 Freiburg Information Technologies GmbH Munzinger Str. 3 REVISIONS Not equipped in Version : 43367260 Revision-No Index Date / Name SCHEMATIC 38803279 060448 23.06.98/MSG Date / Name 063764 22.02.00/MSG 17.12.97/BNH DRAWN 074451 13.08.03/TYR...
Page 50
Marquette Hellige GmbH 388 032 79 - D05 D-79007 Freiburg REVISIONS Not equipped in Version : 43367260 Revision-No Index Date / Name SCHEMATIC 38803279 063764 22.02.00/MSG Date / Name DRAWN 18.12.97/BNH 18.12.97/BNH APPROVED RESPONDER 1000/1100 ISSUED HAGER [1997, 12, 8, 2] /home/projects_b/f67260/design/control...
Page 51
388 032 79 - D06 D-79007 Freiburg REVISIONS Not equipped in Version : 43367260 Revision-No Index Date / Name SCHEMATIC 38803279 063764 22.02.00/MSG Date / Name 067395 05.07.01/TYR DRAWN 18.12.97/BNH 18.12.97/BNH APPROVED RESPONDER 1000/1100 ISSUED HAGER [1997, 12, 2, 3] /home/projects_b/f67260/design/output...
Page 52
388 032 79 - D07 D-79007 Freiburg REVISIONS Not equipped in Version : 43367260 Revision-No Index Date / Name SCHEMATIC 38803279 060448 23.06.98/MSG Date / Name 063764 22.02.00/MSG DRAWN 18.12.97/BNH 18.12.97/BNH APPROVED RESPONDER 1000/1100 ISSUED B. HAGER [1997, 12, 11, 5] /home/projects_b/f67260/design/sync...
Page 55
Marquette Hellige GmbH Sheet: 38803280-D04 D-79007 Freiburg REVISIONS Not equipped in Version : F67261 Revision-No Index Date / Name SCHEMATIC 38803280 07.04.98/MNS Date / Name 15.05.98/MSG DRAWN 19.01.98/MSG 065687 19.10.00/MSG APPROVED 19.01.98/JFS BATTERY MODULE SCHULER ISSUED [1998, 1, 13, 3] /home/projects_b/f67261/design...
Page 60
GE Medical Systems GE Medical Systems Asia Headquarters: Information Technologies, GmbH Information Technologies, Inc. GE (China) Co., Ltd. Munzinger Straße 3-5 No1 Huatuo Road, 8200 West Tower Avenue Milwaukee, WI 53233 USA D-79111 Freiburg Zhangjiang Hi-Tech Park Pudong, Germany Shanghai, P.R.China 201203...