Autosampler Torque - Thermo Scientific EASY-nLC 1200 Troubleshooting And Maintenance Manual

Table of Contents

Advertisement

3
Maintenance Scripts and Service Records
Maintenance Scripts

Autosampler Torque

48
EASY-nLC 1200 Troubleshooting & Maintenance Guide
Use this script to test for a blockage in the solvent system. Before running the script, connect
the Column Out line to the Waste In line as shown in
Figure 34
shows the parameters for the Back Pressure script.
Figure 34. Back Pressure script parameters
❖ To run the Back Pressure script
1. Open the Back Pressure script as follows:
a. Press Maintenance > Scripts.
b. In the Category list, select Test.
c. In the Name list, select Back Pressure.
2. Set up the parameters for this script as follows:
a. Press the Parameters tab.
b. Test solvent A, solvent B, or both A and B by selecting the corresponding check box
or check boxes.
3. Press Start.
The Autosampler Torque script measures the torque required to move the XYZ robot on each
of its axes for the ASC autosampler.
Figure 35
shows the Description page of the Autosampler Torque script.
Figure 32
on
page
45.