Pump Torque

Back Pressure

Thermo Scientific
For information about troubleshooting system leaks, see
System Leak Test"
on
page 183
page
177.
For the pump module in the EASY-nLC 1200 instrument, the Pump Torque script measures
the actuator's ability to generate pressure at predefined torque levels.
Figure 33
shows the parameters for the Pump Torque script.
Figure 33. Pump Torque script parameters
The Back Pressure script determines instrument backpressure for solvent A and B. The script
runs at a preset flow and measures the backpressure on the system.
IMPORTANT Before you perform the Back Pressure script, ensure that the solvent A
bottle contains UHPLC/MS-grade water and the solvent B bottle contains
UHPLC/MS-grade acetonitrile/water (max. 95% acetonitrile). This test is not valid for
other solvents.
3
and
"Troubleshooting a Pump That Fails the Leaks Script"
EASY-nLC 1200 Troubleshooting & Maintenance Guide
Maintenance Scripts and Service Records
Maintenance Scripts
"Troubleshooting the Results of the
on
47