Using the Direct Controls for Troubleshooting and Maintenance

Using the Autosampler Controls

172
EASY-nLC 1200 Troubleshooting & Maintenance Guide
You can control the XYZ robot and the cooler temperature from the Home > Overview page.
❖ To open the Autosampler dialog box
Press the autosampler icon on the Home > Overview page.
The Autosampler dialog box opens (see
Figure 141. Autosampler dialog box
❖ To start the calibration procedure for the current plate format
1. Press Goto Calibration.
The Tools view for the autosampler on the Maintenance > Devices page opens.
2. To calibrate the XYZ robot for the plate format, follow the instructions in
Plates"
on
page
236.
❖ To move the needle to a vial or well position
1. In the Vial or Well box, enter a vial or well position.
2. Press Goto.
❖ To move the needle to a wash bottle position
Press W1, W2, W3, or W4.
Figure
141).
Vial or well position box
Temperature input box
Temperature readback box
"Calibrating
Thermo Scientific