6. Pull out the cooler, as shown in
Figure 102. Pulling out the cooler
❖ To install the new ASC cooler
1. Insert the new cooler.
Be careful not to damage any of the cables during reassembly.
damaged cable.
Figure 103. Damaged cable
2. Insert and tighten the 2.5 mm hex screw in the top center position using a 2.5 mm ball
driver (see
Figure 101
3. Insert and tighten the two 2.5 mm hex screws at the bottom of the cooler using a 2.5 mm
ball driver (see
Figure 100
4. Plug the P-Bus communication cables into the PCB on the back of the cooler, accessing
the PCB through the back of the instrument.
5. Align the back panel to the back of the instrument.
6. Using a Torx T-10 wrench, tighten the eleven screws that secure the back panel to the
instrument.
7. Reconnect the power cable.
8. Turn on the EASY-nLC instrument.
Figure
102.
on
page
116).
on
page
116).
EASY-nLC 1200 Troubleshooting & Maintenance Guide
6
Field Service Maintenance
Replacing the Autosampler Cooler
Figure 103
shows a
Damaged cable
117