Calibrating the Autosampler's XYZ Robot
Managing Plate Formats
Creating a New Plate Format
232
EASY-nLC 1200 Troubleshooting & Maintenance Guide
Figure 172. Tools view for the autosampler
2. Select a plate format from the Plate list.
3. Press Use.
Tip If you are using a Thermo Scientific application or data system to control your
liquid chromatograph/mass spectrometer (LC/MS) system, the plate selection for the
EASY-nLC instrument in the EASY-nLC Configuration dialog box (Foundation
Instrument Configuration) must match the touch-screen application selection.
❖ To create a new plate format
1. Open the Tools view for the autosampler as follows:
a. Press Maintenance > Devices.
b. In the Devices list, select the autosampler device.
c. Press the Tools tab.
Figure 172
shows the Tools view for the autosampler.
Copy To
Calibrate
button
button
Selected plate format