6. Install the new piston seal as follows:
a. With the spring side of the piston seal facing upward, mount the new piston seal onto
the piston seal tool
Figure 58. Piston seal mounted on the piston seal tool
b. Using a pipette, fill the rim of the piston seal with 100% UHPLC/MS-grade
methanol to remove the air from the piston seal spring
Figure 59. Filling the piston seal with methanol
c. Place the pump head on the guide rods and evenly press it all the way down against
the piston seal tool to insert the seal into the pump head
Figure 60. Inserting the piston seal into the pump head
(Figure
58).
Piston seal spring
5
Routine Maintenance
Maintaining the Syringe Pumps
(Figure
59).
(Figure
60).