Returning the EASY-nLC Instrument

Transport Instructions

Thermo Scientific
If it is necessary to return the EASY-nLC instrument to the factory, follow the instructions in
this chapter.
Contents
Transport Instructions
Before packaging the instrument for transport, purge pump A, B and S with
UHPLC/MS-grade methanol. Take care to remove all solvents from the instrument by
running the last purge iteration on air.
❖ To prepare the instrument for transport
1. Press Home > Overview.
2. Press Eject/Insert Tray.
3. Remove the plates/tubes.
4. Remove the waste beaker and the wash/waste bottles.
5. Press Eject/Insert Tray.
6. Close down the system (see
7. Remove the cables on the back side.
8. Fill out the Declaration of Contamination.
9. Place the EASY-nLC instrument in the original shipping container.
"Closing Down the EASY-nLC Instrument"
EASY-nLC 1200 Troubleshooting & Maintenance Guide
10
on
page
22).