Locating Leaks In Or Between Valve S And Valve W - Thermo Scientific EASY-nLC 1200 Troubleshooting And Maintenance Manual

Table of Contents

Advertisement

7
Troubleshooting
Troubleshooting the Results of the System Leak Test

Locating Leaks In or Between Valve S and Valve W

Figure 149. Workflow for locating leaks in the solvent path between valves S and W or in these valves (Part 1)
Valve W might be leaking. Please
contact product support.
188
EASY-nLC 1200 Troubleshooting & Maintenance Guide
If the system leak test returns the following message, follow the workflow in
Figure 150
or the procedure on pages
Check from valve S to valve W, including valve W, for leaks.
System leak test fails and returns the following message:
Check from valve S to valve W, including valve W, for leaks.
Redo the nanoViper fitting connections to the leak test union
Start the Analytical Column Equilibration script
No
Done
Tighten the Viper blind plug in port 1 of valve W
No
Done
Set valve W to the Center position
No
179–196
to locate and remedy the leak.
(step 22
Readback
for flow sensor A
> 100 nL/min
Yes
(step 3a
Readback
for flow sensor A
> 100 nL/min
Yes
(step 3a
on
page
Readback
for flow sensor A
> 100 nL/min
Yes
Go to
Figure 150
on the next page.
Figure 149
(step
11)
on
page
190).
on
page
191).
200).
and
Thermo Scientific