Thermo Scientific EASY-nLC 1200 Troubleshooting And Maintenance Manual page 63

Table of Contents

Advertisement

Thermo Scientific
Figure 28. Control menu
The Acquire Data dialog box opens (see
c. In the Start Mode area, select the Contact Closure option.
Figure 29. Acquire Data dialog box with the Contact Closure option selected
d. Click Start.
The mass spectrometer remains in the Waiting for Contact Closure state until it
receives a contact closure signal.
5. Send the contact closure signal from the EASY-nLC instrument to the MS detector as
follows:
a. On the EASY-nLC touch screen, press Maintenance > Scripts.
The Maintenance Scripts page opens.
b. In the Category list, select Test.
c. In the Name list, select MS Connection.
d. Press Start.
3
Maintenance Scripts and Service Records
Figure
29).
EASY-nLC 1200 Troubleshooting & Maintenance Guide
Maintenance Scripts
41