Thermo Scientific EASY-nLC 1200 Troubleshooting And Maintenance Manual page 284

Table of Contents

Advertisement

A
Error Codes
Table 30. Error codes (Sheet 6 of 8)
No.
Issue code name
69
ERROR_VALVE_SET_
POSITION
70
ERROR_PARALLEL_
PORT_WRITE
71
ERROR_PARALLEL_
PORT_NOT_
INITIALIZED
72
ERROR_PARALLEL_
PORT_MASS_SPEC_
BUSY
73
ERROR_PARALLEL_
PORT_MASS_SPEC_
TIMEOUT
74
ERROR_PARALLEL_
PORT_INIT
75
ERROR_PARALLEL_
PORT_WRONG_PORT
77
ERROR_DEVICE_LOW
_24_VOLT_SUPPLY
78
ERROR_
MANIPULATOR_INIT
79
ERROR_DEVICE_SELF
_TEST
80
ERROR_DEVICE_IN_
BOOT_MODE
262
EASY-nLC 1200 Troubleshooting & Maintenance Guide
Description
COMPONENT_ID
setPosition failed:
Unrecognized position
POSITION/
DEVICE: Low voltage
(VALUE V)
DEVICE: Init failed (not
ready within 1 minute)
DEVICE: Status bit(s) set
after self test:
0xBITS_IN_HEX
DEVICE: Device in boot
mode after device reset
Cause
Valve unable to switch
position
Contact closure port
IO error
Contact closure port
IO error
Contact closure port
IO error
Contact closure port
IO error
Contact closure port
IO error
Contact closure port
IO error
[See issue description.]
[See issue description.]
Error returned from
hardware component
firmware
Error returned from
hardware component
firmware because a
PCB is damaged
Solution
Clean the rotor seal and the
stator. Replace the rotor
seal if necessary.
If the problem persists,
please contact product
support.
Contact Thermo Fisher
Scientific.
Contact Thermo Fisher
Scientific.
Contact Thermo Fisher
Scientific.
Contact Thermo Fisher
Scientific.
Contact Thermo Fisher
Scientific.
Contact Thermo Fisher
Scientific.
Faulty 24V supply.
Connect to remote support.
Autosampler hardware
issue. Turn off the
EASY-nLC instrument, and
turn it back on. If the
problem persists, check the
P-bus cables to the
autosampler.
Connect the system to the
remote support server (see
Chapter
9).
Contact Thermo Fisher
Scientific.