Replacing The Main Power Fuse - Thermo Scientific EASY-nLC 1200 Troubleshooting And Maintenance Manual

Table of Contents

Advertisement

5
Routine Maintenance
Using Only UHPLC/MS-Grade Solvents

Replacing the Main Power Fuse

68
Troubleshooting & Maintenance Guide
Figure 49.
By clicking on the button, you jump to the flow sensor calibration script.
After the flow sensor calibration script has been successfully finished, the "Flow sensor
calibration due" reminder disappears.
For more details see
Flow Sensors
The EASY-nLC uses these fuses:
• For 120 Vac, the instrument uses one T 5 AL 250 V fuse (5 × 20 mm, IEC 60127).
• For 230 Vac, the instrument uses one T 2.5 AL 250 V fuse.
Use only UL Listed and CSA-certified fuses, All fuses supplied with the instrument are UL
Listed and CSA certified.
CAUTION Before removing the fuses, turn off the instrument and remove the power
cable.