Maintenance Scripts

Analytical Column Equilibration

36
EASY-nLC 1200 Troubleshooting & Maintenance Guide
The Analytical Column Equilibration script equilibrates the analytical column with solvent
from pump A.
Use this script to equilibrate or flush the analytical column, determine a suitable equilibration
flow rate for the method, or initialize a new analytical column.
Figure 24
shows the parameters for the Analytical Column Equilibration script.
Figure 24. Analytical Column Equilibration script parameters
Note The Analytical Column Equilibration script sets the valves to these positions:
• Valve A: 1–6
• Valve B: Center
• Valve S: 1–2
• Valve W: 1–2
❖ To run the Analytical Column Equilibration script
1. Open the Analytical Col Equilibration script as follows:
a. Press Maintenance > Scripts.
b. In the Category list, select Prepare.
c. In the Name list, select Analytical Col Equilibration (abbreviated to fit in the space
allotted).