Note Acetonitrile concentrations greater than 95% reduce the lifetime of system
components. An acetonitrile concentration of 80% in water is recommended for solvent
B.
Figure 39
shows the parameters for the Flow Sensors script.
Figure 39. Flow Sensors script parameters
❖ To run the Flow Sensors script
1. Open the Flow Sensors script as follows:
a. Press Maintenance > Scripts.
b. In the Category list, select Calibrate.
c. In the Name list, select Flow Sensors.
2. Set up the parameters for this script as follows:
a. Press the Parameters tab.
b. Select the check boxes for flow sensors A, B, or both that you want to either inspect
or calibrate and inspect.
Note
To calibrate and inspect the flow sensors, this script performs a two-point
calibration as follows:
3
Maintenance Scripts and Service Records
EASY-nLC 1200 Troubleshooting & Maintenance Guide
Maintenance Scripts
53