Using the Direct Controls for Troubleshooting and Maintenance
Using the Valve Controls
Thermo Scientific
You can access the direct controls for the instrument's hardware components by pressing the
component icons on the Home > Overview page.
To use the direct controls, follow these procedures:
•
Using the Valve Controls
•
Using the Pump Controls
•
Using the Autosampler Controls
You can change the positions of the instrument's three-position valves from the
Home > Overview page. For information about the valve positions, see
Pods"
on
page
11.
❖ To change the position of a valve
1. On the Home > Overview page, press the valve icon for the valve that you want to
control.
The Valve dialog box opens (see
position.
Figure 138. Valve dialog box
2. Press the button for the position that you want the valve to switch to: 1–2, Center,
or 1–6.
The readback box displays the new position.
Using the Direct Controls for Troubleshooting and Maintenance
Figure
138). The readback box lists the current valve
Readback box
EASY-nLC 1200 Troubleshooting & Maintenance Guide
7
Troubleshooting
"Six-Port Rotary Valve
165