Thermo Scientific EASY-nLC 1200 Troubleshooting And Maintenance Manual page 283

Table of Contents

Advertisement

Table 30. Error codes (Sheet 5 of 8)
No.
Issue code name
59
ERROR_
MANIPULATOR_SET_
VIAL_CONFIG
60
ERROR_
MANIPULATOR_TRAY
_EJECT
61
ERROR_
MANIPULATOR_TRAY
_LOAD
62
ERROR_
MANIPULATOR_GRID
_ADDRESS
63
ERROR_
MANIPULATOR_
WASH_ADDRESS
64
ERROR_
MANIPULATOR_VIAL_
ADDRESS
65
ERROR_
MANIPULATOR_
NEEDLE_DOWN
66
ERROR_
MANIPULATOR_
NEEDLE_UP
Thermo Scientific
Description
Vial configuration change
failed
(AUTOSAMPLER_RET
URN_CODE)
tray eject failed
(AUTOSAMPLER_RET
URN_CODE)
tray load failed
(AUTOSAMPLER_RET
URN_CODE)
grid address failed
(AUTOSAMPLER_RET
URN_CODE)
wash address failed
(AUTOSAMPLER_RET
URN_CODE)
vial address failed
(AUTOSAMPLER_RET
URN_CODE)
needle down failed
(AUTOSAMPLER_RET
URN_CODE)
needle up failed
(AUTOSAMPLER_
RETURN_CODE)
Cause
Error while saving
autosampler vial
calibration (see code 57
for asr codes)
See code 57 for asr
codes.
See code 57 for asr
codes.
Could not move to
sample position (see
code 57 for asr codes).
Could not move to
wash position (see code
57 for asr codes).
Could not move to vial
position (see code 57
for asr codes).
Could not move needle
down (see code 57 for
asr codes).
Could not move needle
up (see code 57 for asr
codes).
EASY-nLC 1200 Troubleshooting & Maintenance Guide
A
Error Codes
Solution
Calibrated position not
recognized by the
touch-screen application.
See
Chapter 8, "Calibrating
the Autosampler's XYZ
Robot."
Verify that the correct plate
is configured in the
EASY-nLC touch-screen
application and the
EASY-nLC Configuration
dialog box (Foundation
Instrument Configuration).
Check that no objects in
the tray compartment are
obstructing the XYZ robot
movement. Reset
manipulator and run the
Torque script for the ASC
model autosampler.
(See above.)
(See above.)
(See above.)
(See above.)
(See above.)
(See above.)
261