Thermo Scientific EASY-nLC 1200 Troubleshooting And Maintenance Manual page 193

Table of Contents

Advertisement

Thermo Scientific
b. Press W4.
The autosampler needle moves to the W4 position.
c. Press Needle Down.
The autosampler needle descends into the W4 bottle.
d. Press Close to close the Autosampler dialog box.
3. Dispense solvent from pump S as follows:
a. Press the pump S icon.
The Pump S dialog box opens.
b. In the flow rate box, enter the flow rate: 0.01 to 300 µL/min.
c. In the volume box, enter the volume that you want the pump to empty to waste.
d. Press Start.
The pump empties until it pumps the requested volume or reaches the empty
position, whichever occurs first.
Using the Direct Controls for Troubleshooting and Maintenance
EASY-nLC 1200 Troubleshooting & Maintenance Guide
7
Troubleshooting
171