Thermo Scientific EASY-nLC 1200 Troubleshooting And Maintenance Manual page 225

Table of Contents

Advertisement

Figure 159. Solvent A channel blocked at the outlet of flow sensor A
If the flow is <100 nL/min with this
line removed from the solvent path,
install a new solvent line.
Outlet of flow sensor A
plugged with a Viper fitting
Thermo Scientific
d. Check the readback for flow sensor A. Then, depending on the flow rate, do one of
the following:
• If the flow rate is <100 nL/min, do the following:
i.
Stop the Analytical Column Equilibration script.
ii. Disconnect the damaged solvent line from the mixing Tee's inlet for the
solvent A channel.
iii. Install a new solvent line between the mixing Tee's inlet for the solvent A
channel and the flow sensor's outlet.
For ordering information, see
iv. Reconnect the solvent lines, and rerun the system leak test (see
System Leak Test"
• If the flow rate is >100 nL/min, go to the next step.
Troubleshooting the Results of the System Leak Test
Table 32
on
page
on
page
178).
EASY-nLC 1200 Troubleshooting & Maintenance Guide
7
Troubleshooting
268.