Calibrating the Autosampler's XYZ Robot
Replacing the Autosampler Adapter Plate

Replacing the Adapter Plate in the ASC Model Autosampler

230
EASY-nLC 1200 Troubleshooting & Maintenance Guide
Follow the procedure for the autosampler model in your EASY-nLC system to switch between
the two adapter plates as necessary.
"Replacing the Adapter Plate in the ASC Model Autosampler,"
❖ To replace the adapter plate in the ASC model autosampler
1. Press Home > Overview.
The Overview page of the Home menu opens.
2. Make sure that the area in front of the tray compartment is clear, and press Eject/Insert
Tray.
The tray compartment opens.
tray holder. The four studs on the bottom side of the plate fit snugly into the mounting
holes on the tray holder.
Figure 170. Open tray compartment
3. Remove the current adapter plate by pulling it up and away from the tray holder
(see
Figure
171).
Figure 170
shows the vial adapter plate mounted to the
next procedure