Weak or Absent Signal—Leak Upstream of Flow Sensor B
Thermo Scientific
Figure 125
shows a typical pump B profile for a system with a leak upstream of flow sensor B.
Flow sensor B displays a negative flow during the column equilibration and sample loading
steps.
If the gradient is short, a leak upstream of flow sensor B will cause a complete lack of signal
because solvent B is not reaching the columns within the duration of the gradient.
Figure 125. Typical pump B profile for a leak upstream of flow sensor B
Negative flow during the
column equilibration and
sample loading steps
EASY-nLC 1200 Troubleshooting & Maintenance Guide
7
Troubleshooting
Troubleshooting Tips
147