Thermo Scientific EASY-nLC 1200 Troubleshooting And Maintenance Manual page 211

Table of Contents

Advertisement

Figure 150. Workflow for locating leaks in the solvent path between valves S and W or in these valves (Part 2)
Install a new Waste In line.
Then, rerun the system leak test
(page
178).
Reconnect the solvent lines and
rerun the system leak test to
confirm that the system is leak
tight
(page
178).
Install a new Column Out line.
Then, rerun the system leak test
(page
178).
Thermo Scientific
Redo the nanoViper connection to port 2 of valve W
No
Done
Disconnect the Waste In line from the leak test union
Then, block the free end of the union with a Viper fitting.
No
Redo the nanoViper connection to port 3 of valve S
No
Disconnect the Column Out line from port 3 of valve S
Then, block port 3 of valve S with a Viper fitting.
No
Service valve S
(page
(Continued from
Figure
149.)
(step 5
Readback
for flow sensor A
> 100 nL/min
Yes
(step 77
Readback
for flow sensor A
> 100 nL/min
Yes
(step 88
Readback
for flow sensor A
> 100 nL/min
Yes
(step 9•
Readback
for flow sensor A
> 100 nL/min
Yes
82). Then, rerun the system leak test.
EASY-nLC 1200 Troubleshooting & Maintenance Guide
7
Troubleshooting the Results of the System Leak Test
on
page
192)
on
page
193).
on
page
194).
on
page
195).
Troubleshooting
189