Thermo Scientific EASY-nLC 1200 Troubleshooting And Maintenance Manual page 160

Table of Contents

Advertisement

7
Troubleshooting
Troubleshooting Tips
138
EASY-nLC 1200 Troubleshooting & Maintenance Guide
Figure 117. Pressure dip begins later when the system has introduced swept volume
Backpressure is proportional to the viscosity of the solvent mixture. If you are not running a
water/acetonitrile gradient, the pressure profile of your gradient run will differ.
Table 31
on
page 265
lists the viscosities for two-solvent mobile phases consisting of a
water/methanol mixture or a water/acetonitrile mixture.
The pressure starts
to drop at approximately
27 minutes as
100% solvent B
reaches the column.
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents