Calibrating the Autosampler's XYZ Robot
Calibrating Plates
238
EASY-nLC 1200 Troubleshooting & Maintenance Guide
d. In the Plate area of the Tools view for the autosampler, press Calibrate
(see
Figure 174
on
The Autosampler Plate Calibration dialog box opens.
Figure 175
shows the dialog box at the beginning of the plate calibration routine.
The top left and bottom right positions are highlighted in green. The position
readouts for the xyz axes are unpopulated,
Figure 175. Plate view in the Autosampler Plate Calibration dialog box
(3)
3. Press the green vial or well in the upper left corner of the plate view (1).
The needle moves to the current saved coordinate. The saved coordinates for the x and
y axis appear on the left side of the slash, the needle does not descend into the vial or well,
and the z-axis value on the left side of the slash remains at 0.
The values on the left side of the slash change as you adjust the needle position.
page
237).
(1)
(4)
.
(5)
(2)
Thermo Scientific