nanoViper
Figure 2
shows the nanoViper tubing connections in the EASY-nLC 1200 instrument.
Tubing
• Port 6 of valve A to flow sensor A inlet
Connections
• Port 6 of valve B to flow sensor B inlet
• Flow sensor A outlet to mixing Tee inlet
• Flow sensor B outlet to mixing Tee inlet
• Mixing Tee outlet to port 4 of valve S
• Sample loop connected to ports 2 and 5 of valve S
• Column Out line connected to port 3 of valve S and the HPLC union
• Waste In line connected to port 2 of valve W and the HPLC union or the venting Tee
• Autosampler needle connected to port 1 of Valve S