Replacing the Monitor
106
EASY-nLC 1200 Troubleshooting & Maintenance Guide
❖ To remove the monitor
1. Close down the EASY-nLC instrument (see
on
page
22) and unplug the power cable.
2. To remove the top-front panel from the EASY-nLC 1200 instrument, do the following:
a. Using a Torx T-10 wrench, remove the two screws that secure the top-front panel to
the top back of the instrument.
b. Open the tray compartment door, and then use a Torx T-10 wrench to remove the
screw that secures the top-front panel to the front of the instrument.
3. Using a 2 mm hex wrench, remove the four screws that secure the monitor to the
instrument (see
Figure
CAUTION Use only hand tools with a hex tip to avoid damaging the screw heads.
Figure 87. Removing the screws from the front of the monitor
4. Using a #2 Phillips head screwdriver, make a quarter-turn to loosen the three captive
screws that secure the left side panel to the instrument housing. Then remove the panel.
"Closing Down the EASY-nLC Instrument"
87).
Thermo Scientific