Troubleshooting
Troubleshooting Tips

Sample Signal Weak or Absent

Table 17. Possible causes of a weak or absent signal
Symptom
The sample signal is missing or
weak.
144
EASY-nLC 1200 Troubleshooting & Maintenance Guide
If the sample signal is missing or weak, review the troubleshooting tips in
Possible cause
Sample not aspirated
(See
Figure 123
Damaged or incorrect column
(See
Figure 124
Leak in the system
(See
Figure 125
Incorrect column configuration Check the column configuration.
Incorrect solvent in W3 bottle
Incorrect sample loop
configuration
Action
Run the Sample Pickup script. If the Sample
on
page
145.)
Pickup script fails, see
Autosampler Aspiration"
Make sure that you have installed the correct
on
page
146.)
column or columns.
Check the backpressure on the columns by
running the column equilibration scripts.
Run the Leaks test on the system.
on
page
147.)
If the test fails, see
of the System Leak Test"
1. Press Maintenance > Devices.
2. In the Devices list, select EASY-nLC
(HPLC).
3. Press the Properties tab.
4. Clear the One Column Setup check box for
a two-column setup or select this check box
for a one-column setup.
For more information, refer to the EASY-nLC
Series Getting Started Guide.
Make sure that the W3 bottle contains solvent A.
Check the loop configuration on the Properties
page for the EASY-nLC HPLC device on the
Maintenance > Devices page.
Make sure that the configured loop volume
matches the installed sample loop.
Table
17.
"Troubleshooting the
on
page
220.
"Troubleshooting the Results
on
page
183.