Field Service Maintenance
Replacing a Pump
122
EASY-nLC 1200 Troubleshooting & Maintenance Guide
5. Using a #2 Phillips head screwdriver, make a quarter-turn to loosen the three captive
screws that secure the right side panel to the instrument housing. Then remove the panel.
6. Disconnect the stainless steel Viper tubing from the pump and the pressure sensor. This
prevents damage to the fitting of the pressure sensor, because it does not have to be bent
out of the way.
Figure
Figure 108. Fitting on the pump head
108.
Fitting on the pump head