Field Service Maintenance
Replacing the Autosampler
114
EASY-nLC 1200 Troubleshooting & Maintenance Guide
❖ To install the autosampler
1. Insert the new autosampler into the housing.
2. Using a 3 mm L-hex wrench and, if necessary, a needle-nosed pliers or tweezers, insert
and tighten the screws that secure the autosampler to the bottom of the housing as
follows:
• Insert and tighten the screw on the back left side of the autosampler
(see
Figure 96
on
• Insert and tighten the screw on the right side of the autosampler
(see
Figure 95
on
• Insert and tighten the screw on the front of the autosampler
(see
Figure 97
on
3. Insert and tighten the screws that secure the back horizontal cable bar to the autosampler
frame (see
Figure 93
4. Install a new tie wrap to secure the cables to the vertical part of the autosampler frame.
5. Plug the bus cables into the autosampler cooler and horizontal XYZ-axis PCBs.
6. Reinstall the computer box as described in
page
104.
7. Reinsert the autosampler needle into the holder and tighten the fitting.
8. Mount the right side panel to the instrument housing. Then, with a #2 screwdriver,
secure the panel by rotating the three captive screws a quarter-turn clockwise.
Tip Reconnect the left side panel to the housing after you recalibrate the autosampler.
9. Reconnect the power cable.
10. Turn on the EASY-nLC instrument.
11. Calibrate the autosampler as described in
page
113).
page
112).
page
113).
on
page
111).
"To install the new computer box"
Chapter
8.
on
Thermo Scientific