EASY-nLC 1200 Troubleshooting & Maintenance Guide
Closing down the instrument in a controlled manner is important to allow all of the
components to shut down in an orderly sequence. Using this controlled method saves
important data so that the instrument starts with the correct information the next time
you use it.
IMPORTANT With the exception of an emergency shutdown, closing down the
instrument in a controlled manner is important to allow all the components to shut down
in an orderly sequence. If you turn off the power switch during normal operation, you risk
damaging essential system components. Follow the procedure outlined here whenever
possible.
❖ To close down the EASY-nLC instrument and turn off the power
1. Press Exit (lower-left corner of the screen).
A confirmation dialog box opens (see
Figure 19. Confirmation dialog box that opens after pressing Exit
2. Press Power Down.
The EASY-nLC application displays a white screen with a small progress bar. When the
progress bar is filled out and the message appears indicating you can safely turn off the
instrument, go to the next step.
3. After receiving the message that you can safely turn off the instrument, turn off the power
switch on the back panel of the instrument.