Field Service Maintenance
Replacing the Built-In Computer
Thermo Scientific
This chapter provides procedures for use by Thermo Fisher Scientific field service engineers
on replacing the subassemblies in the EASY-nLC instrument. Only Thermo Fisher Scientific
field service engineers can perform these procedures.
CAUTION Only Thermo Fisher Scientific field service engineers can perform the
procedures in this chapter.
Contents
•
Replacing the Built-In Computer
•
Replacing the Monitor
•
Replacing the Autosampler
•
Replacing the Autosampler Cooler
•
Replacing a Rotary Valve
•
Replacing a Pump
•
Replacing a Pump PCB
Replacing the computer box requires the following part and tools.
Tools
• #2 Phillips screwdriver
• Torx T-10 wrench
• Small flathead screwdriver
To replace the built-in computer, follow these procedures:
1.
"To remove the computer box,"
2.
"To install the new computer box"
❖ To remove the computer box
1. Close down the EASY-nLC instrument (see
on
page
22) and then unplug the power cable.
Parts
Computer box:
• P/N LC607 (shipped with Hard drive HDD)
next procedure
on
page 104
"Closing Down the EASY-nLC Instrument"
EASY-nLC 1200 Troubleshooting & Maintenance Guide
6
Field Service Maintenance
Replacing the Built-In Computer
101