Field Service Maintenance

Replacing the Built-In Computer

Thermo Scientific
This chapter provides procedures for use by Thermo Fisher Scientific field service engineers
on replacing the subassemblies in the EASY-nLC instrument. Only Thermo Fisher Scientific
field service engineers can perform these procedures.
CAUTION Only Thermo Fisher Scientific field service engineers can perform the
procedures in this chapter.
Contents
Replacing the Built-In Computer
Replacing the Monitor
Replacing the Autosampler
Replacing the Autosampler Cooler
Replacing a Rotary Valve
Replacing a Pump
Replacing a Pump PCB
Replacing the computer box requires the following part and tools.
Tools
• #2 Phillips screwdriver
• Torx T-10 wrench
• Small flathead screwdriver
To replace the built-in computer, follow these procedures:
1.
"To remove the computer box,"
2.
"To install the new computer box"
❖ To remove the computer box
1. Close down the EASY-nLC instrument (see
on
page
22) and then unplug the power cable.
Parts
Computer box:
• P/N LC607 (shipped with Hard drive HDD)
next procedure
on
page 104
"Closing Down the EASY-nLC Instrument"
EASY-nLC 1200 Troubleshooting & Maintenance Guide
6
Field Service Maintenance
Replacing the Built-In Computer
101