Replacing A Pump Pcb - Thermo Scientific EASY-nLC 1200 Troubleshooting And Maintenance Manual

Table of Contents

Advertisement

6
Field Service Maintenance

Replacing a Pump PCB

Replacing a Pump PCB
124
EASY-nLC 1200 Troubleshooting & Maintenance Guide
5. Prime the pump as described in
6. Mount the tubing back on the pump head. Tighten the fitting using a 1/4 in. open-ended
wrench.
7. Draw fresh solvent into the solvent lines and remove air from the pump head as described
in
"Removing Air After Replacing a Piston Seal or a Pump"
8. Depending on the pump you are replacing, do the following:
• When replacing pump A or B, run the pump leak test, and then run the flow sensor
calibration (see
"Leaks"
• When replacing pump S, run the sample pickup test (see
page
42).
Replacing the pump PCB requires these items:
Tools
• #2 Phillips screwdriver
• 2.5 mm hex wrench
• Small flathead screwdriver
To replace the pump PCB on the PLU pump, follow these procedures:
1.
"To remove a pump PCB,"
2.
"To replace a pump PCB"
❖ To remove a pump PCB
1. Close down the EASY-nLC 1200 instrument as described in
EASY-nLC Instrument"
2. Disconnect the power cable from the instrument's back panel.
3. Using a #2 Phillips head screwdriver, make a quarter-turn to loosen the three captive
screws that secure the right side panel to the instrument housing. Then remove the panel.
4. Disconnect the P-Bus communication cables, the flow sensor cable, and the pressure
sensor cable from the pump.
Figure 110
on
page 125
"Priming the Pump"
on
page 44
and
"Constant Pressure"
Parts
Pump PCB, P/N LC590
next procedure
on
page 127
on
page
22.
shows these cables connected to the pump.
on
page
78.
on
page
81.
on
page
50).
"Sample Pickup"
"Closing Down the
Thermo Scientific
on

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents