Thermo Scientific EASY-nLC 1200 Troubleshooting And Maintenance Manual page 132

Table of Contents

Advertisement

6
Field Service Maintenance
Replacing the Autosampler
110
EASY-nLC 1200 Troubleshooting & Maintenance Guide
Figure 90. Needle connected to the needle holder
5. Unplug the bus cables on the left side of the instrument from the autosampler cooler and
horizontal XYZ-axis PCB (see
Figure 91. Bus cables
Autosampler needle
fitting
Figure
91).
Cooler bus cable
XYZ-axis PCB bus
cable
Thermo Scientific