Troubleshooting
Troubleshooting the Results of the System Leak Test

Locating Leaks Between the Flow Sensors and Valve S

196
EASY-nLC 1200 Troubleshooting & Maintenance Guide
When one or more of the connections are leaking in the solvent path between the flow sensor
outlets and the mixing Tee inlets, the system leak test returns this message (see
page
183:
Check solvent lines between flow sensor A and valve S, including the mixing Tee and flow
sensor B, for leaks.
Follow this section to locate and remedy the leak:
"Locating a Leak Between the Flow Sensors and Valve S"
Table 26