Managing Plate Formats

Selecting a Plate Format

Thermo Scientific
Figure 171. Removing the vial adapter plate before installing the MTP adapter plate.
4. Install the other adapter plate in the autosampler.
When the EASY-nLC instrument is initially installed, the touch-screen application includes
plate formats for vials and microtiter plates.
Follow the appropriate procedure to use an existing plate format, create your own plate
format, or delete a plate format that you no longer need.
"Selecting a Plate Format,"
"Creating a New Plate Format"
"Deleting a Plate Format"
❖ To select the plate format that you want to use
1. Open the Tools view for the autosampler as follows:
a. Press Maintenance > Devices.
b. In the Devices list, select the autosampler device.
c. Press the Tools tab.
Figure 172
shows the Tools view for the autosampler.
next procedure
on
page 232
on
page 234
EASY-nLC 1200 Troubleshooting & Maintenance Guide
8
Calibrating the Autosampler's XYZ Robot
Managing Plate Formats
Mounting stud