5. From the back of the instrument, disconnect the cables shown in
valve to be exchanged.
Figure 104. Cables connected to valves
6. Using a 2.5 mm L-hex wrench, remove the four 2.5 mm hex screws that keep the valve in
place inside this compartment.
7. Pull out the valve, as shown in
Figure 105. Pulling out a valve
❖ To install a new rotary valve
1. Using a small flathead screwdriver, set the rotary switch on the PCB of the new valve,
shown in
Figure
106, to the same address as the valve being replaced:
• Valve A=1
Figure
105.
EASY-nLC 1200 Troubleshooting & Maintenance Guide
6
Field Service Maintenance
Replacing a Rotary Valve