Thermo Scientific EASY-nLC 1200 Troubleshooting And Maintenance Manual page 115

Table of Contents

Advertisement

Thermo Scientific
Figure 73. Flow sensors
4. Remove the black cable from the flow sensor
Figure 74. Removing the cable from the flow sensor
5. Using an 2.5 mm hex wrench, remove the two screws that secure the flow sensor to the
bracket
(Figure
75).
5
nanoViper fitting
Inlet end
Outlet end
nanoViper fitting
(Figure
74).
Troubleshooting & Maintenance Guide
Routine Maintenance
Replacing a Flow Sensor
93