❖ To run the Sample Pickup script
1. Load the autosampler with the appropriate samples.
Tip For more information about running this script and testing the accuracy of the
sample pickup function, see
2. Open the Sample Pickup script as follows:
a. Press Maintenance > Scripts.
b. In the Category list, select Test.
c. In the Name list, select Sample Pickup.
3. Set up the parameters for this script as follows:
a. Press the Parameters tab.
b. In the Volume box [µL], enter the volume to be picked up.
c. In the Flow box [µL/min], enter the flow rate that pump S uses to aspirate the
sample. For aqueous samples enter 20 µL/min.
d. In the Position box, specify the position of the sample in the autosampler. See the
Position dialog box in
4. Press Start.
3
"Checking Sample Pickup"
Figure
30.
EASY-nLC 1200 Troubleshooting & Maintenance Guide
Maintenance Scripts and Service Records
Maintenance Scripts
on
page
222.
43