Thermo Scientific EASY-nLC 1200 Troubleshooting And Maintenance Manual page 255

Table of Contents

Advertisement

Thermo Scientific
2. Do one of the following:
• To create a new plate format, go to
• To copy an existing plate format (copies the format selection), go to
3. Create a new plate format as follows:
a. Press New.
The Create New Plate Format dialog box opens
Figure 173. Create New Plate Format dialog box
b. In the Format list, select a plate format: 6 × 8 (vials), 8 × 12 (96-well plates), or
16 × 24 (384-well plates).
c. In the Name box, enter a name for the new plate format.
d. Press Accept.
The new plate name appears in the plate list with the text [uncalibrated] appended on
the right.
e. Go to
"Calibrating Plates"
4. Copy an existing plate format as follows:
a. Select an existing plate format from the Plate list.
b. Press Copy To.
The Copy Plate Format dialog box opens.
c. In the Name box, type a name for the new plate format.
d. Press Accept.
e. Go to
"Calibrating Plates"
8
Calibrating the Autosampler's XYZ Robot
step
3.
(Figure
on
page 236
to calibrate the vial or well positions.
on
page 236
to calibrate the vial or well positions.
EASY-nLC 1200 Troubleshooting & Maintenance Guide
Managing Plate Formats
step
4.
173).
233