Flow Sensors - Thermo Scientific EASY-nLC 1200 Troubleshooting And Maintenance Manual

Table of Contents

Advertisement

3
Maintenance Scripts and Service Records
Maintenance Scripts

Flow Sensors

52
EASY-nLC 1200 Troubleshooting & Maintenance Guide
Figure 38.
The Flow Sensors script calibrates the flow sensors for pumps A and B.
Use this script to do the following:
• Periodically check the accuracy of the flow sensor calibration for your maintenance
records (select the Inspection Only check box).
Note If the last flow sensor calibration dates back more than 6 months, a reminder will be
shown on the home screen.
• Calibrate the flow sensors when you change the solvent composition of either solvent A or
B (clear the Inspection Only check box).
This calibration does not require any additional plumbing.
IMPORTANT Because the flow sensor calibration requires a stable operating temperature,
ensure that the instrument's side panels are installed and that the instrument has been on
for a minimum of 30 minutes.