Thermo Scientific EASY-nLC 1200 Troubleshooting And Maintenance Manual page 129

Table of Contents

Advertisement

Thermo Scientific
5. Remove the USB, LVDS, and monitor cables connected to the computer box, being
careful not to damage the LVDS plug (see
Figure 88. USB, LVDS, and monitor cables
6. Carefully remove the monitor by pulling it straight forward (see
Figure 89. Removing the monitor
❖ To install the new monitor
1. Carefully insert the new monitor into the instrument housing.
2. If you removed the white plastic needle guide on the back of the monitor, reinstall it.
3. If the computer box has a PS/2 keyboard connector on its back panel as shown in
on
page
16, connect the adapter cable that is supplied with the new monitor to the
monitor cable.
4. Connect the USB, LVDS, and monitor cables to the computer box, being careful not to
damage the LVDS plug (see
Figure
88).
Figure 88
on
page
107).
EASY-nLC 1200 Troubleshooting & Maintenance Guide
6
Field Service Maintenance
Replacing the Monitor
USB cable
Monitor cable
Figure
89).
Figure