Thermo Scientific EASY-nLC 1200 Troubleshooting And Maintenance Manual page 134

Table of Contents

Advertisement

6
Field Service Maintenance
Replacing the Autosampler
112
EASY-nLC 1200 Troubleshooting & Maintenance Guide
8. Move the cable bar to the side (this is easier to do if you also remove the bus cable to
valve B). See
Figure
94.
Figure 94. Horizontal cable bar moved aside
Valve B
9. Using a 3 mm L-hex wrench, remove the three screws that secure the autosampler to the
bottom plate of the housing (see
• Remove the screw from the right side of the autosampler (see
Figure 95. Removing the screw from the right side of the autosampler
Back side of the
autosampler
Figure
95,
Figure
96, and
Right side of the
autosampler
Figure
97):
Figure
95).
Thermo Scientific