Troubleshooting
Troubleshooting Tips
142
EASY-nLC 1200 Troubleshooting & Maintenance Guide
When only the flow path between pump B and the mixing Tee has a blockage, the
backpressure on pump B increases and differs from the backpressure on pump A.
Figure 121
shows the typical profile for a blockage in the solvent system between pump B and
the mixing Tee.
Figure 121. Backpressure shown reaching the maximum pressure for pump B only
Maximum pressure
reached for pump B