Thermo Scientific
6. Cut the tie wrap attaching the cables to the vertical part of the autosampler frame, as
shown in
Figure
92.
Figure 92. Cutting the strip attaching the cables to the autosampler frame
7. Using a Torx T-10 screwdriver, remove the screws that secure the back horizontal cable
bar to the autosampler frame (see
Figure 93. Horizontal cable bar after removal of screws
Horizontal cable bar
Figure
93).
Back left screw that secures the
autosampler to the instrument
EASY-nLC 1200 Troubleshooting & Maintenance Guide
6
Field Service Maintenance
Replacing the Autosampler
Bus cable
connected to
valve B
Autosampler
frame
111