Instructions for use KEY Laser III 1243
|

6 Safety checks

6.1 Scope of the tests

6 Safety checks
Note
At least the basic KaVo service kit is required. The safety checks must be carried
out after the initial commissioning, the maintenance work and work on the safety-
relevant modules, and at intervals of 12 months at the latest.
KaVo recommends carrying out the annual maintenance service before the safety
checks.
See also:
2 Safety, Page 11
Please read laser safety section carefully!
6.1 Scope of the tests
Complete the following tests under observance of IEC 601 and IEC 825 and the
professional indemnity association health and safety regulations (ANSI Z 136.1).
The testing is carried out in each case over 100 pulses at 4 Hz/60, 600 mJ; 6 Hz/
60, 600 mJ;10 Hz/60, 600 mJ; 15 Hz/60, 400 mJ; 20 Hz/60, 250 mJ; 25 Hz/60,
100mJ.