Chapter 6 Communication Requirements

1 Network Requirements

1.1 Insite/Network Connection

The preferred Insite connection uses a broadband modem. This connection requires a
dedicated Ethernet Jack (RJ45) that must be located less than 1 meter (3 feet) from the System
Control Cabinet
For complete descriptions of these connectivity solutions, please refer to the Broadband
Solutions catalogue available through your local GEHC sales and service representative.
Connectivity Process and pre-installations checklists are available in the Broadband
Connectivity PIM available through your local GEHC sales and service representative
InSite requires an Internet Address connecting it to the System. This address must be available
before installing the system. A request form has been defined. For more information, please
refer to
IP Addressing Process
Chapter 6 Communication Requirements
or contact your GEMS OLC representative.
NOTICE
The System Control Cabinet comes equipped with a Firewall unit. The hospital
network must be capable of connecting to this firewall. In the case that it cannot
be, please contact GE Healthcare to discuss alternatives.
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
193