Power assistant and scan on battery

Advertisement

Quick Links

LOGIQ™ E10 Series
Power Assistant and Scan on Battery Guide
The battery protects system data for a limited amount of time
in the event of inadvertent AC power loss.
Options:
Power Assistant-
system power is maintained by the battery
when AC power is lost or the power cord is unplugged.
Reduces the start up time for portable exams. The system will
run up to 20 minutes after AC power is disconnected.
Scan on Battery
(Option on LOGIQ E10s only)- allows the
user to scan using battery power (AC unplugged). With a fresh
battery that is fully charged, the LOGIQ E10s stays powered
for approximately 50 minutes.
Note: Detailed battery function and service information is outlined
in the Basic User Manual.
www.gehealthcare.com

Advertisement

loading

Summary of Contents for GE LOGIQ E10 Series

  • Page 1 LOGIQ™ E10 Series Power Assistant and Scan on Battery Guide The battery protects system data for a limited amount of time in the event of inadvertent AC power loss. Options: Power Assistant- system power is maintained by the battery when AC power is lost or the power cord is unplugged. Reduces the start up time for portable exams.
  • Page 2 LOGIQ E1 0 Series: Pow er Assista nt B a ttery Op tion LOGIQ Battery Icon on the back of To Enter Power Assistant (PA)- Remove the the monitor depicts Power power cord from AC power while the system is on. Assistant is available on the system.
  • Page 3 LOGIQ E1 0 Series: Pow er Assista nt B a ttery Op tion LOGIQ Power Assistant Tips and Tricks The system is not designed to boot up while using the battery, it must be plugged into a power source prior to boot up. The system cannot scan or perform post processing while in PA.
  • Page 4 Note: if this icon displays, DO NOT initiate PA. Contact the Technical Approximate Charging time (empty to full) Service Department or local GE representative. Power Assistant battery- 2.5 hours...
  • Page 5 LOGIQ E1 0 s: Sca n on B a ttery Op tion LOGIQ Battery Icon on Starting Scan on Battery- When the power cord is the back of the unplugged from AC power, the system automatically monitor signifying transitions to Scan on Battery mode. Power Assist and Scan on Battery Scan on...
  • Page 6: Battery Charging

    LOGIQ E1 0 s: Sca n on B a ttery Op tion LOGIQ Battery Charging Battery Status Icon in the Title Bar Whether the system is on or off, the battery will charge as long as the system is connected to AC power via the power cord, and the breaker is on.
  • Page 7: Refreshing The Battery

    LOGIQ E10 Series: B a ttery Op tions LOGIQ Refreshing the Battery – helps maintain the battery life and accuracy of the fuel gauge and helps to avoid unexpected shutdowns. Recommended every 6 months. Battery Deterioration Refresh Procedure: When the system detects battery deterioration Plug the AC cord to the wall outlet and turn on the circuit breaker.