UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
7.13 Replacing the Main Power Fuses
Warning:
Avertissement: Avant de remplacer les fusibles, arrêtez l'instrument. Assurez-vous
1. Remove the fuse cartridge, by using a small screwdriver.
2. Replace the fuses.
Warning:
Avertissement:
Description
Fuse, slow blow (5 x 20 mm)
NCS-3500RS: 4 A
NCP-3200RS: 2 A
3. Reinstall the fuse cartridge.
4. Reconnect the power cord to its source. Turn on the module.
Page 216
Turn off the module. Disconnect the power cord from its source.
de bien débrancher le cordon d'alimentation de la source secteur.
Fig. 84: Fuse cartridge
Always install two fuses. Use only the fuses indicated in the
following table.
Installez toujours deux nouveaux fusibles. Utilisez
uniquement les fusibles indiqués ci-dessous.
Part No.
Included in the Fuses Kit, part no. 6820.0026
For information about the kit, see section 11.3
(→ page 243).
Fuse cartridge
Operating Instructions